2.3.1 % DO Calibration Report
Calibration report gives you detailed information for the % saturation. It
includes date & time, 0% and 100% saturation mV value, % saturation
offset calibration and barometric pressure at which the solution was
measured.
2.3.2 To View Calibration Report:
1. From % saturation measurement mode, press left or right arrow
key to navigate to other available functions until you see
REPO
function in the LCD.
2. Press
REPO (F2)
key. The calibration report is shown in the display.
3. Press
PRIN (F2)
to transfer the calibration report to computer through
IrDA.
Figure 54: DO Calibration Report in % Saturation Mode
Function Keys available in saturation report screen:
PRIN (F2)
Transfers calibration report to Computer through IrDA
NEXT (F3)
Goes to saturation measurement mode
ESC (F4)
Goes to saturation measurement mode
ENTER
(Not functional)
(Not functional)