1. Press up & down arrow keys to enter the first digit of the password and
then press
NEXT (F3)
key to move to the next digit.
2. The next digit is selected. Press up & down arrow keys to enter the
second digit of the password. Enter all 5-digits.
3. Press
ENTER
key to confirm the password.
1. When the correct password is entered, the Setup Key Function Screen
appears.
2. Press
Enter
key to launch Setup Selection Screen. Press up or down
arrow key to go to required setup sub-group.
3. Press
ENTER
key to select the sub-group.
2.5 Modifying Setup parameters
1. Press
NEXT (F3)
key to select individual setup parameters sequentially.
2. Press (Up) or (Down) arrow key to change the value of a selected
parameter.
3. Once you have changed a value:
•
Press
ENTER
key to save the change, or
•
Press
NEXT (F3)
key to go to the next parameter without saving the
changed parameter.
4. Press
NEXT-P (F2)
or
PRE-P (F1)
to navigate to next or previous page.
5. Press
ESC (F4)
to exit from setup mode.
(Refer to Page 9 for current function keys settings)
Note:
You can enter ‘00000’ (read-only password) if you wish to view the
setup parameters. You are not allowed to modify any parameter when
you enter ‘read-only password’.
Note
: If you enter an incorrect password, the screen shows “Try Again”. If
an incorrect password is entered three consecutive times, the meter goes
into measurement mode. If you forget the password, there is no way to
access the system setting and calibration. Please contact your authorised
dealer for assistance.
Summary of Contents for CD 650
Page 2: ......
Page 4: ......
Page 10: ...2 ...
Page 28: ...20 ...
Page 45: ...37 ...
Page 49: ...41 ...
Page 51: ...43 ...
Page 61: ...53 ...
Page 65: ...57 ...
Page 67: ...59 ...
Page 73: ...65 ...
Page 82: ...74 END OF CONDUCTIVITY SECTION ...
Page 84: ...76 ...
Page 92: ...84 ...
Page 98: ...90 ...
Page 100: ...92 ...
Page 106: ...98 ...
Page 108: ...100 ...
Page 126: ...118 ...