PART A - % Saturation Mode
1. O2 % - DO Saturation Setup
The DO saturation (%) setup screen presents many options to control the
operating parameters, which can be controlled and set from the DO setup
screen. The settings are displayed in 2 pages. Press
NEXT-P (F2)
and
PREV-P
(F1)
to navigate through these pages.
DO (%) Settings Page 1 – Operating Parameters
Figure 49 : DO (%) – Page 1
Parameter
Function
Factory
Default
Offset Cal
Allows to adjust % saturation offset calibration
0.0 %
Calibration due
Specify number of days for the DO calibration alarm. The
meter shows CAL DUE indicator after calibration date.
Available range : 0 to 30
5 days
Select Pressure Unit
Sets the unit for barometric pressure
Available units: mmHg and kPa
mmHg
Measured Pressure
Barometric pressure as measured by the instrument.
Adjusted Pressure
User adjusted value.
-
Pressure
Compensation
ENABLE – The meter will compensate for the barometric
pressure at the location depending on the altitude.
DISABLE – No compensation for the pressure.
(This is applicable only for the % saturation mode. For mg/L
or ppm mode pressure comp. would always be applicable)
ENABLE
Summary of Contents for CD 650
Page 2: ......
Page 4: ......
Page 10: ...2 ...
Page 28: ...20 ...
Page 45: ...37 ...
Page 49: ...41 ...
Page 51: ...43 ...
Page 61: ...53 ...
Page 65: ...57 ...
Page 67: ...59 ...
Page 73: ...65 ...
Page 82: ...74 END OF CONDUCTIVITY SECTION ...
Page 84: ...76 ...
Page 92: ...84 ...
Page 98: ...90 ...
Page 100: ...92 ...
Page 106: ...98 ...
Page 108: ...100 ...
Page 126: ...118 ...