62
7. The upper display shows the next calibration point.
8. Rinse the ISE with clean water. Place it in the next calibration solution and
follow the steps 4 through 6 to calibrate other points.
9. Press
ESC (F4)
to go to measurement mode.
2.2 Calibration Report
Calibration report gives you detailed information on calibration. It includes date
& time, calibration points (Ion concentrations), absolute mV reading for each
point and slope.
Figure 25 : Ion Calibration Report
Example:
In the given Ion calibration report, calibration has been done for
0.01, 0.1 & 10 Ion concentrations. The corresponding absolute mV readings
are 0.0mV, 56.0mV & 156mV. The slope is 56 mV/decade for the range
enveloped from 0.01 to 0.1 ppm. Similarly, the slope is 50 for the range
enveloped from 0.1 to 10ppm where 0.1 calibration point has been skipped.
2.2.1 To View Calibration Report
1. From ION measurement mode, press left or right arrow key to
navigate to other available functions until you see
REPO
function in the
LCD
2. Press
REPO (F2)
key. The calibration report is shown in the display.
3. Press
PRIN (F2)
to transfer the calibration report to computer through IrDA.
Note:
The function of the keys available on the Ion calibration/ report
screen is same as previously described in pH calibration with a user-
defined buffer section.
Note:
The meter accepts the second and subsequent points only if the mV
readings are within the calibration window (
15mV/decade to 90mV/decade
).
Summary of Contents for CD 650
Page 2: ......
Page 4: ......
Page 10: ...2 ...
Page 28: ...20 ...
Page 45: ...37 ...
Page 49: ...41 ...
Page 51: ...43 ...
Page 61: ...53 ...
Page 65: ...57 ...
Page 67: ...59 ...
Page 73: ...65 ...
Page 82: ...74 END OF CONDUCTIVITY SECTION ...
Page 84: ...76 ...
Page 92: ...84 ...
Page 98: ...90 ...
Page 100: ...92 ...
Page 106: ...98 ...
Page 108: ...100 ...
Page 126: ...118 ...