7-11 No responses when pressing measurement start switch
Main causes in case of no responses when pressing measurement start switch:
・
Failure of measurement start switch
・
Failure of control board
Replace the control board.
Replace the measurement
start switch.
No response when pressing the measurement start switch.
OK
Is conduction between Pin 7 and Pin 8 of
CN 403 of the control board normal when
pressing the measurement start switch?
NO
YES