23
7-12 No responses when pressing operation switch
Main causes in case of no responses when pressing operation switch:
・
When wearing gloves
・
Failure of operation switch
・
Failure of LCD relay board
・
Failure of operation switch board or control board
No response when pressing the operation switches.
Replace the control board.
OK
NO
YES
YES
Is the power voltage between TP301 and
TP302 of the LCD relay board 3.3V?
NO
Is the power voltage between Pin3 and
Pin8 of CN306 of the LCD relay board
changed from 3.3V to 0V when
pressing the leftmost operation switch?
Replace the LCD cover unit.
Do you wear gloves?
YES
NO
Replace the LCD
relay board.
Take off your gloves.