26
7-15 Cannot see target
Main causes when you cannot see target
・
When power saving mode is ON
・
Failure of control board
・
Failure of optical unit
You cannot see the target.
NO
OK
Can you see the target?
NO
Are the optical unit and control
board connected correctly?
Connect the harness correctly.
Replace the optical unit ASSY.
Replace the control board.
YES
YES
Set the power saving mode as OFF.
NO
YES
Is the power saving mode ON?