Chapter 6 Checks and Maintenance
82
1. Open the cabinet door.
2. Lift the protective grille outwards from the bottom and then upwards
to detach.
②
①
Fig. 6.5 How to detach the protective grille
3. Clean the surfaces of the humidifying tray and humidifying heater with
a brush or by other means.
4. Reattach the grille and close the cabinet door.
Test Area Cleaning
Dirt and foreign matter inside the test area can throw test results off.
Clean the test area before every test.
1. Open the cabinet door.
2. Wipe walls and parts clean with a soft cloth.
3. Close the door.
Water Circuit (Heat Exhaust) Compartment Cleaning
Dust buildup inside the water circuit (heat exhaust) compartment can
lead to trouble. Clean inside the compartment once a month.
1. Check the main power switch is in the OFF position.
The water circuit (heat exhaust) compartment has a ventilation fan on
the inside. Contact would be dangerous. Always shut OFF power at
the main power switch before opening the water circuit (heat exhaust)
compartment.
Procedure
Procedure
Procedure