Chapter 5 Getting More Out of Your Cabinet
72
11. Dry-bulb temperature calibration
Stands for
“
Calibration Temperature
”
.
Amount of offset
:
Change the value.
:
Selects the digit.
ENT.
:
Enters the selection.
* The target temperature can be set within the below range.
-5.0
≦
Calibration temperature
≦
5.0
○
SYSTEM
•
With the calibration feature, you are basically setting the
number of degrees from the current test area
temperature that must be added or subtracted in order to
display the temperature you want. For example, if the
target temperature is 60.0°C and the calibration value is
changed from 0.0°C to –0.2°C while testing is in progress,
the display will read “59.8°C.” After that, test area
temperature is controlled to 60.0°C, or 0.2°C higher than
what the target temperature was prior to setting the
calibration value.
•
How to read the current dry-bulb temperature
Press the
key until the far right digit starts flashing.
At that moment, the dry-bulb temperature (test area
temperature) will be displayed in the message area for 1
to 2 seconds.
Dry-bulb
(test area)
temperature
○
SYSTEM
Calibration
temperature
Press the
NEXT
key and move to the next setting.