Chapter 4 Operation
56
15. Repeat counter end step
Input “3” and press the
ENT.
key. Then, press the
NEXT
key.
Repeat counter end step
Change the value.
ENT.
Enters the selection.
* The repeat counter end step can be set within the following
range. Repeat counter start step
Repeat counter end
step
Number of registered steps
If the repeat range setup does not satisfy the below condition,
an error will be generated when you attempt to save the
program (“SAVE”
ENT.
key) and you will not be able to
save the program. “EDIT ERR” will appear on the display,
therefore correct the repeat range to satisfy the condition.
Repeat counter start step
Repeat counter end step
Number of registered steps
Stands for “End”.
PRGM SET
Press the
NEXT
key and move to the next setting.
16. End mode HOLD/OFF/CNST
Select “OFF” and press the
ENT.
key. Then, press the
NEXT
key.
End mode
HOLD Holds the conditions created in the last
step.
OFF
Shuts control power OFF, but primary
power remains ON.
CNST
Operates the constant mode conditions.
ENT.
Enters the selection.
Stands for “Last”.
PRGM SET
Press the
NEXT
key and move to the step edit content selection
(step 2 in this procedure).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com