Chapter 4 Operation
50
8. Step exposure time
Input “1.00” and press the
ENT.
key. Then, press the
NEXT
key.
Step exposure time
Change the value.
Selects the digit.
ENT.
Enters the selection.
* Exposure time can be set between 00 (h): 00 (min) and 99
(h): 59 (min), or between 100 and 999 hours.
PRGM SET
h. min.
Press the
NEXT
key and move to the next setting.
9. Soak control status ON/OFF
Select “OFF” and press the
ENT.
key. Then, press the
NEXT
key.
Soak control status
ON
Turns soak control ON. Go to the step data
enter command (step 12 in this procedure).
* In soak control, the exposure time for a
given step is counted down only while the
temperature (humidity) is within the set
attainment range for the set holding time.
So, the time specimens are exposed to the
target temperature will equal the time set
in step 8 above. See “5.2 User Setup and
Safety Device Options Setup”.
OFF
Turns soak control OFF. Go to the
temperature ramp control setting (step 10 in
this procedure).
* The exposure timer will start at the same
time the step does and will continue
counting without stopping.
ENT.
Enters the selection.
Stands for “Guaranteed Soak”.
PRGM SET
* Soak control cannot be used when either temperature
or humidity ramp control is used.
Press the
NEXT
key and move to the next setting.