Chapter 4 Operation
38
4.3 Specimen Safety Device Setup
Overheat Protector
CAUTION
•
Always set the overheat protector BEFORE
beginning tests. Set it according to the type of
specimens and test conditions.
Unless properly set, the overheat protector will not trip in the
event of cabinet trouble, which could lead to specimen damage.
For details on the overheat protector setting, see "1.2 Specimen
Protection (Safety Devices)".
Turn the dials until a suitable temperature is displayed.
OVERHEAT
PROTECTOR
Dial
Fig. 4.14 Overheat protector
Absolute High/Low Limit Temperature (Humidity) Alarms
CAUTION
•
Always set the absolute high/low limit temperature
(humidity) alarms BEFORE beginning tests. Set
alarms according to the type of specimens and test
conditions.
Unless properly set, the absolute high/low limit temperature
(humidity) alarms will not be triggered in the event of cabinet
trouble, which could lead to specimen damage. For details on
alarm settings, see "1.2 Specimen Protection (Safety Devices)".
Alarm Setup Mode
This section explains how to set the temperature alarms that are built
into the temperature controller. Alarms available for setting are the
absolute high limit temperature, absolute low limit temperature (&
humidity) and upper deviation limit temperature alarms.
The absolute high and low limit temperature (& humidity) alarms are
Procedure