ergoline ergoselect 400 Operator'S Manual Download Page 37

- 37 -

Use the softkeys 

↑ ↓

 to select the parameter to edit.

At Select, for example, you can choose the protocol type:

‑ Step (load increase in steps) or
‑ Ramp (continuous load increase).

Press SELECT to save the selected protocol type.

To cancel the selection, press the 

 key.

All other parameters are edited in the same way.

Using the arrow keys (

↑ ↓

), highlight a parameter and 

confirm the selection with SELECT: the corresponding 
value appears in reverse video and can be changed with 
the arrow keys 

↑ ↓

.

Pressing SELECT will save the new value.
You exit the configuration with 

.

C

ontrast

The display contrast is adjustable in the range from 
0 to 100%.

s

electing

 

the

 

paraMeter

 

to

 

edit

Protocol 6.

Select  

Step

Basic Load 

25  Watt

Stage Time 

  2  min

Load Stage 

25  Watt

NIBP Lead Time 60  sec
Recovery Load  25  Watt
Recovery Time    2  min 

Select

Choose function

Protocol 6.

Select

 

 Step

Basic Load 

25  Watt

Stage Time 

  2  min

Load Stage 

25  Watt

NIBP Lead Time 60  sec
Recovery Load  25  Watt
Recovery Time    2  min 

Select

Choose function

e

diting

 

the

 

paraMeter

 

value

e

diting

 

the

 

paraMeter

 

value

Contrast

 

50 %

 

Select

Choose function

Summary of Contents for ergoselect 400

Page 1: ...ergoselect 400 Arm Ergometer Operator s Manual 201000160000 Version 2020 01 14 Rev 03 English...

Page 2: ...d in this manual All trademarks appearing in this document are trademarks of their respective owners Their protection is acknowledged No part of this manual may be reprinted translated or reproduced w...

Page 3: ...atient 15 Operation 16 Control terminal P 16 Turning the System On 16 Operating Modes with Control Terminal P 17 PC Mode 18 Ergometry 19 Manual 21 Settings with Control Terminal P 22 Control Terminal...

Page 4: ...4 4...

Page 5: ...ce inspection and repair of the device must read and understand the content of this manual before using or work on it Paragraphs with special symbols are of particular importance If unauthorized indiv...

Page 6: ...gometer is connected to other equipment or if a medical system is created it must be ensured that the added leakage currents do not present a hazard In case of questions please contact your ERGOLINE d...

Page 7: ...ff pressure remains constant for some time the measurement will also be aborted and the cuff deflated Contraindications The following patient categories are excluded from using the device patients fee...

Page 8: ...ughly instructed on the basis of the instructions for use patients of the intended patient group who have been thoroughly instructed by trained specialists The group of intended users does not include...

Page 9: ...ressure actuation CE mark per the Medical Device Directive 93 42 EEC of the European Union Notified body T V S D Product Service GmbH Ridlerstr 65 80339 M nchen Germany Nationally Recognized Testing L...

Page 10: ...l K 2 Handgrips 3 Special chair 4 Mains connection 5 Power switch green button 6 Cable connections 7 Foot rests 8 Strain relief for cable 9 Adjustable feet to compensate for uneven floor conditions 10...

Page 11: ...r adjustable feet should now evenly protrude from the base plate see illustration Using the nut M10 5 mm thick screw one additional adjustable foot into the guide tube as far as possible do not yet ti...

Page 12: ...rside of the ramp with a fork wrench SW 17 6 You can now slide the support tubing of the ramp onto the guide tube again and the ramp is ready for use Application Lift the wheelchair at the front and c...

Page 13: ...eter to the power line check that the line voltage corresponds to the ratings on the type plate The type plate is located on the back of the ergometer at the bottom Power input and power switch Cautio...

Page 14: ...ower cord and the connection cables Open the clamping lever at the chair s guide tube and take off the chair pulling it towards the rear When you have reached the new location lower the ergometer very...

Page 15: ...s on the control terminal If the optional access ramp for wheelchairs is used posi tion the front wheels on the ramp and lock the wheelchair Check that all clamping levers are tightened Wheelchair loc...

Page 16: ...5 keys With this key you display the main menu or return to the previous menu level With this key you initiate a blood pressure measure ment A measurement in progress can be aborted with the same key...

Page 17: ...oad changes via the keypad SETTINGS Used to configure the ergometer Speed readout At the top of the control terminal there is a speed readout for the patient as well as three LEDs that inform the pa t...

Page 18: ...d the corresponding values will be displayed The exercise test can only be terminated with the corre sponding command from the controlling ECG unit Note All functions are locked while the ergometer is...

Page 19: ...ction with SELECT The exercise test is started with the Start key a blood pressure measurement at rest may precede the test see Settings When the basic load appears on the display after approx 15 seco...

Page 20: ...ed manually It is recommended that the patient continues to pedal in the recovery phase The END key in the middle will terminate the test Note The saddle height ergoselect 200 can be changed during an...

Page 21: ...measurements The exercise test is started with the Start key afterwards the load can be set and changed with the 5 W and 5 W keys in increments of 1 W up to 25 W as configured Blood pressure measureme...

Page 22: ...will display this menu Use the softkeys on the right and left to position the bar cursor on your preferred default mode and save the selection with SELECT Protocols Protocols 6 15 are user programmab...

Page 23: ...w keys Pressing SELECT will save the new value You exit the configuration with Contrast The display contrast is adjustable in the range from 0 to 100 Load Change Here you determine the increments for...

Page 24: ...in with you select DATE or TIME and confirm the selection Then the value displayed in reverse video can be edited with the keys and saved with SELECT The time is adjusted in the same way You exit the...

Page 25: ...be initiated with digital commands Digital default The communication with the ergo meter is entirely controlled with digital commands Analog IN OUT The entire communication load control and BP mea su...

Page 26: ...e RPM limit values RPM Min 0 70 54 Max 50 130 64 Select Note The limits selected in this menu only apply to the load range between 6 and 150 watts At higher loads the RPM limits automatically adapt to...

Page 27: ...be aborted with the same key The functions of these six softkeys change with the displayed menu the key label describing the function is shown on the display Keypad K Exercise Test PC Mode Training M...

Page 28: ...ns Speed readout 1 speed low patient should pedal faster 2 correct speed 3 speed high patient should pedal slower Note If during an exercise test the speed drops below 30 RPM the load readout starts b...

Page 29: ...only be terminated with the corre sponding command from the controlling ECG unit Note All functions are locked while the ergometer is operating in PC mode except for the saddle height adjustment and t...

Page 30: ...rate are represented both by numeric values and waveforms When the basic load appears on the display after approx 15 seconds or upon termination of the blood pressure measurement and the patient s RP...

Page 31: ...overy phase terminate the test Pressing again displays another menu where you can change the saddle height and the display mode see PC Mode Terminating the Test Once the full protocol has been complet...

Page 32: ...t and changed with the Load and Load keys in increments of 1 W up to 25 W as config ured Blood pressure measurements can be initiated with Terminating an Exercise Test The exercise test can be termina...

Page 33: ...t rate are represented both by numeric values and by waveforms When the basic load appears on the display after approx 15 seconds or upon termination of the blood pressure measurement and the patient...

Page 34: ...rotocols are saved to the chip card by means of a PC program ergoline opticare professional or ergo line opticare basic Upon completion of the training session the entire proce dure incl load and hear...

Page 35: ...raining session automatic ter mination when the programmed recovery phase has been completed or manual termination the test subject can state how the test was perceived BORG scale Subsequently all tra...

Page 36: ...ault mode and save the selection with SELECT Protocols Protocols 6 15 are user programmable protocols 1 5 are fixed see Appendix for protocol parameter details Standard values for the following parame...

Page 37: ...nd can be changed with the arrow keys Pressing SELECT will save the new value You exit the configuration with Contrast The display contrast is adjustable in the range from 0 to 100 Selecting the param...

Page 38: ...ice each key press will change the load by 1 5 10 und 25 watts Language The texts can be displayed in different languages Beep The audio signal emitted during blood pressure measure ments can be turne...

Page 39: ...mode pulse constant interval there will be different parameters to define for the training phase First of all you select and confirm the protocol you wish to configure Then you select the parameters...

Page 40: ...modes pulse constant load and interval the warmup phase the duration of the training session and the recovery phase are defined first Depending on the selected training mode you can edit the correspo...

Page 41: ...tiated with digital commands Digital default The communication with the ergometer is entirely controlled with digital commands Analog IN OUT The entire communication load control and BP mea surements...

Page 42: ...RPM limit values Note The limits selected in this menu only apply to the load range between 6 and 150 watts At higher loads the RPM limits automatically adapt to the respective loads Load watts Green...

Page 43: ...mage and or discolor the upholstery Disinfection Only the following disinfectants are approved for disinfec tion Sch lke Mayr GmbH Antifect AF FF FD 10 Terralin 0 5 Quartamon Med B Braun Melsungen AG...

Page 44: ...ng alcohol Clean the microphone with a cloth moistened with alcohol or soap water Allow the microphone to dry before reinserting it in its pocket Disinfection For disinfection spray a disinfectant spa...

Page 45: ...ce Engineer authorized by ergoline Similarly the automatic sphygmomanometer in the control terminal must be checked and calibrated by an authorized specialist every two years to fulfill legal requirem...

Page 46: ...rogrammable Internal Protocols Control Terminal P 5 fixed incremental exercise test protocols e g WHO 10 user programmable protocols manual load control Control Terminal K 5 fixed incremental exercise...

Page 47: ...method oscillometric for resting BP the results from both measurements are compared for plausibility Measuring Range systolic pressure 40 to 280 mmHg diastolic pressure 40 to 280 mmHg pulse rate 35 to...

Page 48: ...25 99 Adjustment Range 20 100 1 30 1 400 20 100 1 99 With Control Terminal P the recovery load is fixed at 25 W Test Protocols Control Terminal K only Protocol initial load W duration sec load increme...

Page 49: ...nge black speed independent range to DIN VDE 0750 0238 black grey speed independent range of the ergoselect ergometer Family of characteristics of the load periods according to IEC 60601 1 Under perma...

Page 50: ...peration in the configuration in which it is being used Guidance and Manufacturer s Declaration Electromagnetic Emissions The ergoselect ergometer is intended for use in the electromagnetic environmen...

Page 51: ...differential mode 2 kV common mode 1 kV N A Mains power should be that of a typical commercial or hospital environment Voltage dips short interrup tions and voltage variations on power supply input l...

Page 52: ...om fixed RF transmitters as deter mined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marke...

Page 53: ...t Power of Transmitter W Separation Distance According to Frequency of Transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0...

Page 54: ...54...

Page 55: ......

Page 56: ...ergoline GmbH Lindenstra e 5 72475 Bitz Germany Tel 49 0 7431 98 94 0 Fax 49 0 7431 98 94 128 e mail info ergoline com http www ergoline com...

Reviews: