![ergoline ergoselect 400 Operator'S Manual Download Page 11](http://html1.mh-extra.com/html/ergoline/ergoselect-400/ergoselect-400_operators-manual_2422184011.webp)
- 11 -
e
rGometer
s
etuP
The ergoselect 400 is mounted on a wooden pallet for
shipment.
You need a wrench (SW 17) or a ratchet and the corre‑
sponding socket to detach the ergometer from the pallet
and for all other mounting operations described below.
m
ountInG
the
s
PeCIal
C
haIr
• Take the shipping box off the ergoselect 400 and
remove the packaging accessories from the box.
• Detach the chair from the pallet and remove.
• Using the fork wrench SW17, unscrew the 4 screws on
the underside of the pallet.
• Using the nut M10 (8 mm thick), screw the 4 adjust‑
able feet to the underside of the base plate, observing
an 8‑mm distance between the nut M10 and the
hexagon head of the adjustable foot.
Tighten each nut M10.
All four adjustable feet should now evenly protrude
from the base plate. (see illustration)
• Using the nut M10 (5 mm thick), screw one additional
adjustable foot into the guide tube as far as possible ‑
do not yet tighten the nut M10.
• Lift the base plate carefully from the pallet and place
it on a horizontal, level ground.
• It should now be easy to slide the chair onto the guide
tube. If necessary, the height of the guide tube with
respect to the chair can be modified with the four ad‑
justable feet. If the ground is even, all four feet must
be adjusted by the same height.
• Now unscrew the adjustable foot of the chair's guide
tube until it is flush on the floor, but make sure that
the foot does not lift the ergoselect 400 base plate off
the ground.
Tighten the nut M10 with a fork wrench SW 17.
a
djusting
the
feet
Caution
• Equipment Damage •
NEVER lift the device by the chair's guide tube!
a
djustable
foot
of
the
guide
tube
Summary of Contents for ergoselect 400
Page 1: ...ergoselect 400 Arm Ergometer Operator s Manual 201000160000 Version 2020 01 14 Rev 03 English...
Page 4: ...4 4...
Page 54: ...54...
Page 55: ......