![ergoline ergoselect 400 Operator'S Manual Download Page 25](http://html1.mh-extra.com/html/ergoline/ergoselect-400/ergoselect-400_operators-manual_2422184025.webp)
- 25 -
eKG t
yPe
The selected EKG Type determines the communication
method with the ECG recorder, PC‑based ECG system, etc.
To prevent an accidental change of this setting, the menu
is protected with a password.
Using the arrow keys, enter 003 and confirm the entry
with SELECT.
All ergoselect ergometers support the following communi‑
cation modes:
• Analog with pulse
Remote start mode; prior the each load change, the
ergometer generates a control pulse and sends the
corresponding data via the interface.
• Analog / Digital
An analog voltage controls the load ‑ blood pressure
measurements can be initiated with digital commands.
• Digital (default)
The communication with the ergo meter is entirely
controlled with digital commands.
• Analog IN‑OUT
The entire communication (load control and BP mea‑
surements) is controlled with analog signals.
No digital data will be sent.
Select the communication mode and confirm with SELECT.
s
electing
the
ergoMeter
coMMunication
Mode
EKG Type
Analog with pulse
Analog / Digital
Digital
Analog IN-OUT
Select
Note
• The EKG Type needs to be selected only when the ergo-
meter is connected to an ECG unit. The selection is part
of the installation procedure.
• The "Analog/Digital" and "Digital" communication is only
possible when PC Mode is selected from the main menu
or when this is the default mode.
e
ntering
the
eKg t
ype
password
EKG Type
003
Select
Summary of Contents for ergoselect 400