- 53 -
G
UIDANCE
AND
M
ANUFACTURER
’
S
D
ECLARATION
– E
LECTROMAGNETIC
I
MMUNITY
The ergoselect ergometer is intended for use in the electromagnetic environment specifi ed below. It is the responsibility of
the customer or user to ensure that the ergoselect ergometer is used in such an environment.
ImmunityTest
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment - Guid-
ance
Electrostatic discharge (ESD)
to EN 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV
± 8 kV
Floors should be wood, concrete or ceramic tile.
If fl oors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast transient/burst
to EN 61000-4-4
± 2 kV for power supply lines
± 1 kV for input and output lines
± 2 kV passed
Mains power should be that of a typical commer-
cial or hospital environment.
Surge to EN 61000-4-5
± 1 kV diff erential mode
± 2 kV common mode
± 1 kV
N/A
Mains power should be that of a typical commer-
cial or hospital environment.
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines to
EN 61000-4-11
< 5 % UT
(> 95 % dip in UT)
for 0.5 cycles
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
< 5 % UT
(> 95 % dip in UT) for 5 s
< 5 % UT
40 % UT
70 % UT
< 5 % UT
Mains power should be that of a typical commer-
cial or hospital environment. If the user of the
ergoselect ergometer requires continued operation
during power mains interruptions, it is recommend-
ed that the ergoselect ergometer be powered from
an uninterruptible power supply or a battery.
Power frequency (50/60 Hz)
magnetic fi eld to
EN 61000-4-8
3 A/m
passed
Power frequency magnetic fi elds should be at levels
characteristic of a typical location in a typical
commercial or hospital environment. The ergoselect
ergometer has no components susceptible to
magnetic fi elds.
NOTE: UT is the a.c. mains voltage prior to application of the test level.