- 19 -
O
PERATING
M
ODES
WITH
C
ONTROL
T
ERMINAL
P
An ergoselect ergometer with a control terminal P supports
the following operating modes:
PC MODE
An external device (e.g. stand-alone electrocardio-
graph, PC-based ECG system) controls the ergometer -
no intervention at all is required at the ergometer.
ERGOMETRY
The ergometer runs an automatic exercise test - some
of the corresponding test protocols are user-confi gu-
rable and stored in the system (see chapter "Settings").
MANUAL
The ergometer is controlled manually, i.e., the user
performs all load changes via the keypad.
SETTINGS
Used to confi gure the ergometer.
S
PEED
READOUT
At the top of the control terminal, there is a speed readout
for the patient as well as three LEDs that inform the pa tient
of the speed: too slow, too fast or correct.
The ranges for the respective speed ratings depend on the
selected load (see "Technical Specifi cations").
S
PEED
READOUT
1
speed low (patient should pedal faster)
2 correct
speed
3
speed high (= patient should pedal slower)
Note
•
If, during an exercise test, the speed drops below 30 RPM,
the load readout starts blinking on the display.
•
To reactivate the saddle height adjustment function, press
and the arrow keys will again be displayed.
•
Additional blood pressure measurements an be initiated
with
.