- 50 -
Infl ation Pressure
300 mmHg max.; during infl ation the infl ation pressure
automatically adapts to patient's BP
Infl ation Rate
between approx. 6 seconds (to 140 mmHg) and
approx. 18 seconds (to 300 mmHg)
Max. Cuff Pressure
300 mmHg
Cuff Defl ation Method
pulse-dependent defl ation rate
approx. 3 mmHg/beat or approx. 3 mmHg/s
Calibration
calibration with external pressure meter
Artifact Rejection
automatic artifact rejection and comparison of the rest-
ing BP readings from both methods for plausibility
E
XERCISE
T
EST
P
ROTOCOLS
Protocol
initial load
[W]
time in stage
[min]
load
increment
[W]
recovery load
[W]
recovery time
[min]
1. WHO
25
2
25
25
99
2. BAL
50
3
50
25
99
3. Hollmann
30
3
40
25
99
4. STD France
30
3
30
25
99
5. Standard
20
1
25
25
99
6. - 15. (user-programmable)
25
2
25
25
99
Adjustment Range
20 - 100
1 - 30
1 - 400
20 - 100 (*)
1 - 99
(*) With Control Terminal P, the recovery load is fi xed at 25 W.
T
EST
P
ROTOCOLS
(C
ONTROL
T
ERMINAL
K
ONLY
)
Protocol
initial load
[W]
duration
[sec]
load
increment
[W]
time in stage
[sec]
recovery load
[W]
recovery time
[min]
ramping protocol
0
120
25
10
25
99
PWC-130
(*)
25
0
25
120
25
99
PWC-150
(*)
50
0
25
120
25
99
PWC-170
(*)
50
0
50
120
25
99
(*) The program advances to the recovery phase as soon as the target heart rate (130/150/170) is reached.