ELite 5, User Manual
Page 50 of 159
Figure 312. Prepare for shipment
If the ‘ELite 5’ analyzer is powered off without a proper shutdown or prepare
for shipment procedure, the following may occur:
•
The pneumatic components may lock in an undefined state. The next
startup and pneumatic initialization may take longer with increased
reagent consumption.
•
The ‘ELite 5’ application software and/or the Windows
®
XP
®
Embedded operating system can suffer damage or data loss.
•
The ‘ELite 5’ analyzer may have to be restarted by a service engineer
after an improper shut down or power off sequence.
The ‘Prepare for Shipment’ procedure should be used for following reasons:
•
The reagents may spill out if the ‘ELite 5’ is tilted, potentially causing electrical short
circuit or corrosion.
•
Reagents can flow out of the ‘ELite 5’ if the reagent intakes are disconnected.
•
Evaporation may cause reagents left in the pneumatic components in the ‘ELite 5’
analyzer to become concentrated and form crystals that increase blank results,
damage pneumatic components, and may require a service engineer to restore
proper operation.
Summary of Contents for ELite 5
Page 13: ...ELite 5 User Manual Page 13 of 159 Figure 1 ELite 5 with Autosampler Space Requirements...
Page 27: ...ELite 5 User Manual Page 27 of 159 Figure 9 4DIFF Scatter Diagram...
Page 142: ...ELite 5 User Manual Page 142 of 159 20 3 Printed Report Formats...
Page 143: ...ELite 5 User Manual Page 143 of 159...
Page 153: ...ELite 5 User Manual Page 153 of 159 20 5 Fluidic System...