ELite 5, User Manual
Page 93 of 159
10 Calibration
The ‘ELite 5’ analyzer must be calibrated in order to provide correct, reliable results.
Calibration should be done at the following times:
•
At initial installation,
•
After replacing any component critical to the dilution or measurement process,
•
Any time quality control shows a systematic error or is outside predefined limits,
•
At periodic time intervals determined by laboratory regulations.
The ‘ELite 5’ analyzer calibration process consists of running a commercial calibration
material or a human whole blood sample with known parameter values multiple times. The
known parameter values and the average values of multiple runs are used to calculate
calibration factors. The new calibration factors take effect once they are accepted by the
operator.
If you are using a human whole blood as a calibrator, the calibration target values must be
measured on a reference analyzer according to applicable regulations, standards, or
laboratory procedures.
An automated calibration procedure is provided that guides the operator through the
calibration process and performs the calibration factor calculation. Calibration factors can
also be modified directly without running a calibration procedure. Only the parameters that
are traceable to reference methods are calibrated: WBC, RBC, PLT, HGB, MCV and MPV.
To view the calibration functions of the ‘ELite 5’ analyzer, select the ‘Calibration’ icon on the
main menu to display the calibration panel. The Calibration panel displays an option to
calibrate the ‘ELite 5’ analyzer or to view prior calibrations.
Figure 66. Calibration Options
Summary of Contents for ELite 5
Page 13: ...ELite 5 User Manual Page 13 of 159 Figure 1 ELite 5 with Autosampler Space Requirements...
Page 27: ...ELite 5 User Manual Page 27 of 159 Figure 9 4DIFF Scatter Diagram...
Page 142: ...ELite 5 User Manual Page 142 of 159 20 3 Printed Report Formats...
Page 143: ...ELite 5 User Manual Page 143 of 159...
Page 153: ...ELite 5 User Manual Page 153 of 159 20 5 Fluidic System...