ELite 5, User Manual
Page 137 of 159
18.7 Dilutor Errors
•
Pinched, clogged tube around dilutor or at the rear reagent tubes
•
Physical obstruction (foreign material)
18.8 Priming Problems
18.8.1 The Analyzer Would Not Prime Liquids
•
The respective reagent is out
•
Aspirating tube (in container)
-
has fallen off,
-
has a leakage,
-
is broken,
•
A reagent puffer is leaking… look for liquid below the analyzer
•
Damaged tube in the system… look for leakage, or traces of liquid, contact
service
18.9 Electronics Related Problems
18.9.1 No Image on Display, No Backlight
•
Contact Service
18.9.2 Touch Sensitive Surface Not Working
•
Connect an external USB mouse
o
this will allow clicking on the screen elements
o
contact service for resolution of the problem
18.9.3 Touch (Click) Is Inaccurate
•
The touch screen needs calibration
•
Connect an external USB mouse
o
perform the calibration under Main Menu / Maintenance / Touch screen
calibration
18.9.4 The Cursor Seems To Be Moving With Good Ratios, But In a Smaller Area
•
Try to calibrate the touch screen (use a mouse to access the functions)
•
If the problem persists, contact service
18.10 The Analyzer Does Not Power On
•
Check proper power connections, and the state of the (small) switch at the
power supply
•
If the above fails, contact service
18.11 I2c Errors Displayed At Startup
•
Contact Service
Summary of Contents for ELite 5
Page 13: ...ELite 5 User Manual Page 13 of 159 Figure 1 ELite 5 with Autosampler Space Requirements...
Page 27: ...ELite 5 User Manual Page 27 of 159 Figure 9 4DIFF Scatter Diagram...
Page 142: ...ELite 5 User Manual Page 142 of 159 20 3 Printed Report Formats...
Page 143: ...ELite 5 User Manual Page 143 of 159...
Page 153: ...ELite 5 User Manual Page 153 of 159 20 5 Fluidic System...