ELite 5, User Manual
Page 119 of 159
The information panel also contains a Collect button. By pressing this, there is a chance to
save a .gz file to the selected path. This file can be sent to the local service engineer. The file
contains valuable information about the instrument including self test, stress measurement
and the calibration factors.
If the instrument has a problem, the collect file will help the engineer to analyze the problem,
check the measurements and read the log file.
With this function the service engineer has a chance to study the instrument from a distance
and give valuable information to the end user.
Before pushing the collect button, the user has to plug in a pen drive to make the system
able to find this external storage after collect button is pressed. After the pen drive is found,
the user can browse it to save the .gz file.
Summary of Contents for ELite 5
Page 13: ...ELite 5 User Manual Page 13 of 159 Figure 1 ELite 5 with Autosampler Space Requirements...
Page 27: ...ELite 5 User Manual Page 27 of 159 Figure 9 4DIFF Scatter Diagram...
Page 142: ...ELite 5 User Manual Page 142 of 159 20 3 Printed Report Formats...
Page 143: ...ELite 5 User Manual Page 143 of 159...
Page 153: ...ELite 5 User Manual Page 153 of 159 20 5 Fluidic System...