TopCal S CPC 300
6 Commissioning
Hauser
35
T7
Glass el. 7.0
Glass el.4.6
Antimony
IsFET
Select electrode type
(only pH)
!
Note!
• In the event of a change from glass
or antimony electrode to IsFET, the
temperature sensor is reset to Pt 1000 as
a default. Conversely, Pt 100 is selected.
• Dependent on the ordered measuring
input, "Glass el. 7.0" or "IsFET" is dis-
played as default value.
T9
solution ground
no solution ground
Select connection type
solution ground
= with potential matching
(PML)
no solution ground
= without PML
T10
°C
°F
Select temperature display
T11
ATC
MTC
MTC+Temp
Select temperature compensation
ATC
= automatic temperature
compensation
MTC
= manual temp. comp. (with fixed
temperature, entered in Field GAA2)
MTC+Temp
. = as MTC. On the display
however, the value which appears is what
the temperature sensor measures in the
medium.
T12
025.0 °C
Temperature value
(only with pH and selection of MTC or
MTC+Temp. in Field T11)
T13
off
on
Pt 100
Pt 1000
NTC 30k
Temperature measurement
(only for redox)
CODE
Display
CHOICE
(default = bold)
INFO
Hold
pH
7.00
Param
Edit
E
Glas El. 7.0
Glas El. 4.6
Antimon
IsFET
pH electr.typeK1
Next
Hold
E
Edit
pH
7.00
Param
Sensor ground
solution ground
no solution ground
Hold
pH
7.00
Param
Edit
E
°C
°F
Temp. unit
pH
7.00
Param
Edit
ATC K1
MTC
MTC+Temp
E
Temp.conf. K1
Next
Hold
pH
7.00
Param
Edit
-20.0...150.0°C
E
0
25.0
MTC-Temp. K1
Next
Hold
mV -
114
Param
Edit
E
Temp.meas1
Next
on
off