First commissioning page 33
TopCal S CPC 300
2
EHauser
First commissioning .................................... page 33
Using the menu "Quick Setup" in the chapter "First Start-up" you can configure the
transmitter with the most important parameters quickly and easily. Important basic
functions like language, measuring parameter, temperature compensation and current
outputs are configured with the help of the on-site display.
You carry out further configuration in the corresponding menus as required.
Brief operating instructions
Change electrode
Remove electrode.
•
Move service switch into
position.
•
Note error message.
•
Remove electrode.
Install electrode.
•
Service switch must be in
position.
•
Note error message.
•
Install electrode.
•
Switch service switch to
.
•
Note item in display.
Manual operation
•
Move assembly.
•
Start programme.
•
Stop programme.
"PARAM": Parameter entry menu
"PARAM" allows you to return to the previous "return field" from any point in the menu.
"DIAG": brings you to the instrument diagnosis menu.
Help page: Press "DIAG" and "PARAM" together
"MEAS": Operation
Press "MEAS" to exit any of the menus ("PARAM", "DIAG", "CAL") without having com-
pleted the settings / calibration.
"CAL": Calibration key for manual electrode calibration.
"E": (Enter): Continue in menu / confirm your selection
LED: "green" = everything OK. "red" = error.
Arrow keys:
• Scroll through menu options and highlight a selection or
• increment/decrement numbers by one step with "+" / "-".
Next digit: using the "right arrow" (editor type 1) or
• "Activate" with the "right arrow" and scroll through the selection with "+" / "-"
(editor type 2).
PARAM