6 Commissioning
TopCal S CPC 300
42
EHauser
!
Note!
•
The controller function "contiuous controller" can only be assigned to the current
output 2.
•
Note on table entry: Enter the number of support points (value pairs) in Field EAB1.
Enter the value pairs in Field EAB2. For example. (4 support points):
ƒig. 25:
Example of characteristic enter using a table
linear:
EAA1
0/4 mA:
02.00 pH
/ 000.0 °C / -0500 mV
20 mA:
12.00 pH
/ 100.0 °C / 0500 mV
Entry of the upper and lower measured value limits
The maximum range of the measured value is
–2 ... +16 pH.
The minimum distance from the upper to the lower meas-
ured value limit is 2 pH units.
(Exmp.: 0/4 mA: pH 7 and 20 mA: pH 9)
EAA6
Linear characteristic active.
Note in display:
The linear characteristic is activated after confirmation
by pressing "E". Cancel by pressing "PARAM".
EAB1
01
(1 ... 10)
Entry of the number of support points (value pairs)
EAB2
pH/Redox/°C/:
000.0
mA:
04.00
Value pair entry
pH/Redox/°C - mA (number of required value pairs =
number of support points desired in Field EAB1).
EAB3
OK
Delete element(s)
Selection:
Are the value pairs OK or do you want to delete
elements?
EAB4
pH/Redox/°C/:
000.0
mA:
04.00
Delete:
Select the rows to be deleted, delete them with
T
and
confirm this with "E".
EAB5
Valid table
Note in display (no entry)
Table status If invalid, then back to EAB4.
EAB6
Table active
Note in display:
The table is active after confirmation by pressing "E".
Cancel by pressing "PARAM".
CODE
CHOICE
(default = bold)
INFO
Current output1
(Terminals 31 +, 32 –)
Current output2
(Terminals 33 +, 34 –)
pH/redox
Temperature
pH/redox
Temperature
contiuous controller
8
4
2
6 8
12
20
mA
16
pH
C07-CPM153xx-05-06-00-xx-002.EPS