10 Technical data
RMS 621
58
Hauser
10.0.2
Output
Output signal
Current, pulse, transmitter power supply and switching output
Galvanic isolation
Basic device:
!
Note!
The specified insulation voltage is the AC testing voltage U
eff.
, which is applied between the
connections.
Basis for assessment: EN 61010-1, protection class II, over voltage category II
Current - pulse output variable
Current
•
0/4 to 20 mA +10% overreach, invertible
•
Max. loop current 22 mA (short-circuit current)
•
Max. load 750
Ω
at 20 mA
•
Accuracy 0.1% of full-scale value
•
Temperature drift: 0.1% / 10 K ambient temperature
•
Output ripple < 10 mV at 500
Ω
for frequencies < 50 kHz
•
Resolution 13 Bit
•
Error signals 3.6 mA und 21 mA limit as per NAMUR NE43
Pulse
Basic device:
•
Frequency range 0.5 Hz to 12.5 kHz
•
Voltage level 0 to 1 V low, 24 V high ±15%
•
Min. load 1 k
Ω
•
Max. pulse width 100 ms for frequencies < 4 Hz
Expansion cards (digital passive, open collector):
•
Frequency range of 0.5 Hz to 12.5 kHz
•
I
max.
= 200 mA
•
U
max.
= 24 V ± 15%
•
U
low/max.
= 1.3 V at 200 mA
•
Max. pulse width 100 ms for frequencies < 4 Hz
Connection,
terminals
Power
supply
(L/N)
Input 1/2
0/4 to 20
mA/PFM/pul
se (10/11) or
(110/11)
Input 1/2
TPS
(82/81) or
(83/81)
Input 1/2
temperature
(1/5/6/2) or
(3/7/8/4)
Output 1/2
0 to 20
mA/pulse
(132/131)
or
(134/133)
Interface
RS232/485
housing
front or
(102/101)
TPS
external
(92/91)
Power supply
2.3 kV
2.3 kV
2.3 kV
2.3 kV
2.3 kV
2.3 kV
Input 1/2
0/4 to 20
mA/PFM/pulse
2.3 kV
500 V
500 V
500 V
500 V
Input 1/2 TPS
2.3 kV
500 V
500 V
500 V
500 V
Input 1/2
temperature
2.3 kV
500 V
500 V
500 V
500 V
500 V
Output 1/2
0 to 20
mA/pulse
2.3 kV
500 V
500 V
500 V
500 V
500 V
500 V
Interface
RS232/RS485
2.3 kV
500 V
500 V
500 V
500 V
500 V
500 V
TPS external
2.3 kV
500 V
500 V
500 V
500 V
500 V
Summary of Contents for RMS621
Page 1: ...BA127R 09 en 08 04 510 04788 Software Version 2 0 Operating manual RMS 621 Energy manager ...
Page 6: ...RMS 621 6 Endress Hauser ...
Page 23: ...RMS 621 5 Operation Endress Hauser 23 5 1 2 Display Quick Setup ...
Page 88: ......
Page 90: ...www endress com worldwide BA 127R 09 en 08 04 510 04788 FM SGML6 0 ...