Proservo NMS5
Hauser
145
Matrix group
Function group
Item
Access
code
Short description
Default value
Set
Select
Display
Possible settings,
selection, or display
Index No,
GVH
OPE. SELECT
840
51
TANK PROFILE
1 min.
Select
841
Set
843
Set
844
846
847
OPE. POINT
I/F MANU. LEVEL
51
1 min.
Set
BAL. LEVEL
OPE. WAIT TIME
51
51
51
O: stop
2
0 mm
2.0 mm
PROFILE OPE.
UP WAIT TIME
LIQ. WAIT TIME
NOTE!: Data in G8V4H0
to G8V4H9 are used for
both Tank and Interface
0ROlLEMEASUREMENT
3ETTHENUMBEROFMEASUREMENTPOINTSFOR4ANK0ROlLE
measurement.
51
Select method for density measurement.
SPOT4ANK0ROlLE)&0ROlLE-!.5)&0ROlLE
51
Manually set the water interface level to be referenced during
4ANK0ROlLEMEASUREMENT
3ETTHEALLOWANCEFORLEVELMOVEMENTPRIORTO4ANK0ROlLE
OPERATION)FLEVELMOVEMENTEXCEEDSTHISVALUE0ROlLEOPERATI
ON
ISCANCELED3ETTINGMMALLOWS0ROlLEMEASUREMENTTO
proceed, regardless of level movement
Set waiting time for displacer weight to be measured in air, when
MAKINGTHEWEIGHTTABLEATSTARTOF0ROlLEOPERATION
Set the time for displacer to stop at each measurement position
DURINGPROlLEOPERATION
E.g. if level movement exceeds setting at G8V4H4, Proservo is in
"stand by" for the time set in this matrix. After the maximum
TIMEELAPSESPROlLEOPERATIONISCANCELED
1 min.
Set
Set
Set
0: stop
TANKPROlLE
)&PROlLE
-!.5)&PROlLE
2 - 16
0 - 99999.9 mm
1.0 - 99.9 mm
1 - 31 min
1 - 31 min
1 - 31 min
845
13 Matrix