Installation
Proline Prowirl 73
10
Hauser
Outlet runs with pressure and temperature measuring points
If pressure and temperature measuring points are installed after the device, please ensure there
is a large enough distance between the device and the measuring point so there are no negative
effects on vortex formation in the sensor.
A0003780
2.2.6 Vibrations
The correct operation of the measuring system is not influenced by plant vibrations up to 1 g,
10 to 500 Hz. Consequently, the sensors require no special measures for attachment!
PT
TT
3...5 x DN
4...8 x DN
Esc
E
-
+