Electromagnetic
compatibility (EMC)
• As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21)
• Device version with PROFIBUS DP: Complies with emission limits for industry as per EN
50170 Volume 2, IEC 61784
The following applies for PROFIBUS DP: If baud rates > 1.5 MBaud, an EMC cable
entry must be used and the cable shield must continue as far as the terminal wherever
possible.
Details are provided in the Declaration of Conformity.
16.9 Process
Medium temperature range
–20 to +150 °C (–4 to +302 °F)
0
0
0
0
-20
-40
-40
20
20
40
100
60
140
[°F] [°C]
40 60 80 100
100
120 140 160
200
300
180
360
-20
-40
-40
T
A
T
F
[°F]
[°C]
A0027806
40
Promag 500 – digital
T
A
Ambient temperature range
T
F
Fluid temperature
Summary of Contents for Proline Promag H 500