Diagnostics and troubleshooting
Proline Promag H 500 PROFIBUS DP
182
Hauser
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
387 HistoROM backup failed
Contact service organization
• Conductivity
• Corrected conductivity
• Measured values 1
• Measured values 2
• Measured values 3
• Density
• Electronic temperature
• Empty pipe detection
• Flow velocity
• Low flow cut off
• Mass flow
• Reference density
• Corrected volume flow
• Temperature
• Volume flow
Measured variable status
Quality
Bad
Quality substatus
Maintenance alarm
Coding (hex)
0x24 to 0x27
Status signal
F
Diagnostic behavior
Alarm
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
512 Sensor electronic (ISEM) faulty
1. Check ECC recovery time
2. Turn off ECC
• Conductivity
• Corrected conductivity
• Density
• Electronic temperature
• Empty pipe detection
• Flow velocity
• Low flow cut off
• Mass flow
• Reference density
• Corrected volume flow
• Temperature
• Volume flow
Measured variable status
Quality
Bad
Quality substatus
Maintenance alarm
Coding (hex)
0x24 to 0x27
Status signal
F
Diagnostic behavior
Alarm
12.7.3 Diagnostic of configuration
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
330 Flash file invalid
1. Update firmware of device
2. Restart device
• Conductivity
• Corrected conductivity
• Measured values 1
• Measured values 2
• Measured values 3
• Density
• Electronic temperature
• Empty pipe detection
• Flow velocity
• Low flow cut off
• Mass flow
• Reference density
• Corrected volume flow
• Temperature
• Volume flow
Measured variable status
Quality
Bad
Quality substatus
Maintenance alarm
Coding (hex)
0x24 to 0x27
Status signal
M
Diagnostic behavior
Warning