Proline Promag H 500 PROFIBUS DP
Diagnostics and troubleshooting
Hauser
185
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
441 Current output 1 to n
1. Check process
2. Check current output settings
–
Measured variable status [from the factory]
Quality
Good
Quality substatus
Function check
Coding (hex)
0xBC to 0xBF
Status signal
S
Diagnostic behavior
Warning
1)
Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
442 Frequency output 1 to n
1. Check process
2. Check frequency output settings
–
Measured variable status [from the factory]
Quality
Good
Quality substatus
Function check
Coding (hex)
0xBC to 0xBF
Status signal
S
Diagnostic behavior
Warning
1)
Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
443 Pulse output 1 to n
1. Check process
2. Check pulse output settings
–
Measured variable status [from the factory]
Quality
Good
Quality substatus
Function check
Coding (hex)
0xBC to 0xBF
Status signal
S
Diagnostic behavior
Warning
1)
Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.