Proline Fieldcheck
Device functions - thermal measuring systems
Hauser
79
Function group PARAMETER
Verification of the amplifier takes place along four measuring points MP1 to MP4. Use this function
group to enter the full scale value MP4 of the verification flow (100%) and the two selectable
measuring points MP2 and MP3 (in % of the full scale value).
!
Note!
The value of MP1 is prespecified (5% of the full scale value MP4).
Function description
FUNCTION
VERIFICATION
PARAMETER
FLOW 100%
Displays the full scale value of the verification flow.
On the basis of the original factory calibration, the value is adapted to the application
settings (medium, pressure, temperature etc.) and cannot be changed for the verification.
Unit:
The unit configured in the measuring device is adopted.
MEAS. POINT 2/3
Use this function to enter the measuring points MP2 and MP3 (in % of the full scale
value).
User input:
MP2: 10 to 90 %
MP3: 15 to 95 %
!
Note!
• The value of MP2 must be less than the value of MP3.
• Both values are stored permanently.
Factory setting:
MP2: 10 %
MP3: 50%