Wiring
Proline Fieldcheck
28
Hauser
3.13
Connecting to Prosonic Flow 92 in single-chamber
housing
"
Caution!
Connect the ground terminals on the cables to the uninsulated housing components or to the
ground terminals in the connection compartment.
1.
Remove the cover of the electronics compartment (5), remove the display module and lift up
the plastic cover.
2.
Disconnect the signal cable connectors (6) from the connection board.
3.
Connect the connectors of the Simubox cable to the adapter (3).
4.
Release the securing screw and fold down the cover (7).
5.
Connect Fieldcheck to the current and frequency outputs of the transmitter (8) (terminal
pairs 22/23, 24/25 or 26/27).
!
Note!
The configuration of the outputs depends on the product type and the order version of the
tested transmitter. To find out which terminal pairs are assigned to the outputs, see the
"Wiring" section of the flowmeter Operating Instructions.
6.
Connect Fieldcheck to the service connector (4).
A0013417
Fig. 21:
Connecting Fieldcheck to Prosonic Flow 92 (Ex i version)
1
Fieldcheck flow simulator
2
Simubox
3
Adapter, order number: 71110558
4
Service connector connection (use Fieldcheck/CDI connecting cable, order number 71032689
5
Electronics compartment cover
6
Signal cable connector
7
Connection compartment cover
8
Connecting cable for Fieldcheck/current and frequency outputs