Proline Fieldcheck
Device functions - electromagnetic measuring systems
Hauser
59
Group VERIFICATION SENSOR
Function description
FUNCTION
VERIFICATION SENSOR
INSTRUMENT
VERSION
Use this function to enter the measuring device version.
Options:
COMPACT / REMOTE:
Remote version with a cable length of up to 20 meters max. or compact version
REMOTE > 20 m:
Remote version with a cable length longer than 20 meters
!
Note!
The HV isolation test is not available for the remote version with a cable length longer
than 20 meters.
Factory setting:
COMPACT / REMOTE:
ISOLATION TEST
Use this function to select whether an extended isolation test (against undesired ground
connections) should be carried out for all signal shields and the mag coil.
Options:
WITH HV TEST PULSE
NO HV TEST PULSE
Factory setting:
WITH HV TEST PULSE
TEST MAGNETIC COIL
Use this function to check the magnetic coil of the Promag sensor. If you select the
W. COIL TEST option, you will be prompted to switch off the measuring device before
verification starts. This is necessary to ensure that the coil connector is safely
disconnected and reconnected.
Options:
W. COIL TEST
NO COIL TEST
Factory setting:
W. COIL TEST
CONDITION PIPE
Use this function to specify the status of the measuring tube.
!
Note!
You are only provided with the possibility of selecting the APPLICAT. REF. DATA
function group if the measuring tube is full on completion of sensor verification
Options:
PIPE ENTIRELY FILLED
PIPE EMPTY
Factory setting:
PIPE ENTIRELY FILLED
APPLICATION
REFERENCE DATA
Use this function to select whether the reference data of the application should be
accepted at the end of the sensor verification process.
Options:
YES
NO
Factory setting:
YES