Function group PARAMETER
Verification of the amplifier takes place along four measuring points MP1 to MP4. Use this function
group to enter the full scale value MP4 of the verification flow (100%) and the two selectable
measuring points MP2 and MP3 (in % of the full scale value).
!
Note!
The value of MP1 is prespecified (5% of the full scale value MP4).
Function description
FUNCTION
VERIFICATION
PARAMETER
FLOW 100%
Use this function to enter the full scale value of the verification flow.
User input:
Max 6-digit floating point number (without sign)
!
Note!
Only values between 1 and 8 m/s are permissible. The input limits MIN and MAX are
displayed below the input line.
Unit:
The unit configured in the measuring device is adopted.
Factory setting:
Equivalent to 4 m/s (depending on the unit)
MEAS. POINT 2/3
Use this function to enter the measuring points MP2 and MP3 (in % of the full scale
value).
User input:
MP2: 10 to 90 %
MP3: 15 to 95 %
!
Note!
• The value of MP2 must be less than the value of MP3.
• Both values are stored permanently.
Factory setting:
MP2: 10 %
MP3: 50%