Device functions - Coriolis measuring systems
Proline Fieldcheck
62
Hauser
11.1.1
Group SIMULATION
Function group CONFIGURATION
Function description
FUNCTION
SIMULATION
CONFIGURATION
OPERATION MODE
Use this function to select the operation mode for the simulation.
There are two freely adjustable flow signals of –12.5 m/s to +12.5 m/s and the assigned
duration of simulation is 1 second to 999 seconds.
Options:
CONTINUOUS
The flow behavior in the process has low dynamics.
a0001708
Q1 = Simulation value 1
STEP
The flow rates in the process are very dynamic and change very quickly. This operation
mode is also suitable for the simulation of flows outside the tolerance limits (fault
simulation). Here very dynamic or disrupted conditions can also be simulated.
a0001709
Q1 = Simulation value 1, Q2 = Simulation value 2, T1 = Duration of simulation 1,
T2 = Duration of simulation 2
Factory setting:
CONTINUOUS
FLOW SPECIFICATION
Use this function to select which measurement is to be simulated.
Options:
MASS FLOW
VOLUME FLOW
CORRECTED VOLUME FLOW (not for Promass 40)
!
Note!
• The CORRECTED VOLUME FLOW option is only available if the FIXED STANDARD
DENSITY option was selected in the CORRECTED VOLUME CALCULATION
function in the tested device.
• For the simulation of corrected volume flow, the system uses the values specified by
the flowmeter in the STANDARD DENSITY and UNIT STANDARD DENSITY
functions.
Factory setting:
MASS FLOW
Q
Q1
START
STOP
t
Q
Q1
Q2
T2
T1
START
STOP t