Commissioning
Micropilot NMR84
62
Hauser
9.4.2
Configuration of a connected Prothermo temperature
transmitter
If a Prothermo NMT532, NMT539 or NMT8x temperature transmitter is connected via
HART, it can be configured as follows:
1. Navigate to: Expert → Input/output → HART devices → HART Device(s) → NMT device
config; here,
HART Device(s)
is the name of the connected Prothermo.
2. Go to the
Configure device?
parameter and select
Yes
.
3.
a
A0030098
27
Prothermo NMT53x: Position of the bottom temperature element
a
Distance from bottom temperature element to zero reference (tank bottom or datum plate).
To configure a
Prothermo NMT53x
: Go to the
Bottom point
parameter and enter
the position of the bottom temperature element (see picture above).
The value entered into the
Bottom point
parameter in the Tank Gauging device
is handed over to the
Bottom point
parameter in the connected
Prothermo NMT53x.