Operating menu
Micropilot NMR84
206
Hauser
The table editor on the local display
1. Navigate to Setup → Advanced setup → Application → Tank calculation → Dip-table
→ Table mode and select the
Disable
option.
2. Navigate to Setup → Advanced setup → Application → Calculation → Dip-table → Edit
table
13953-1
3
2
1
10.0
5.0
1.0
10.1
4.2
1.0
Dip Table Di.
/../Edit table
N Dip Table De.
A0045687
57
The dip table editor on the local display
N
Number of the line
De. Device level
Di. Dip level
3. Use the "↑" and "↓" keys to move to the line you want to edit.
13964-1
5
4
3
1.0
0.0
0.0
1.0
0.0
0.0
Dip Table Di.
Dip Table De.
/../Edit table
N
A0045873
4. Press "E" to open the line.
5. Use "→" to select the cell you want to edit.
13964-1
5
4
3
1.0
0.0
0.0
1.0
0.0
0.0
/../Edit table
N
Dip Table Di.
Dip Table De.
A0045874
6. Press "E" to open the cell.
7. Enter the required number → 44.
8. Continue until all required table points have been entered.
9. Press "-" and "+" simultaneously to quit the table editor.
10. Navigate to Setup → Advanced setup → Application → Tank calculation → Dip-table
→ Table settings and select the
Sort table
option.
The table points are arranged in an ascending order.
11. Navigate to Setup → Advanced setup → Application → Tank calculation → Dip-table
→ Table modeand select the
Enable
option.
The new dip table is active.