Operating menu
Micropilot NMR84
114
Hauser
Additional information
Read access
Operator
Write access
Maintenance
Air density
Navigation
Operation → Density → Air density
Description
Defines the density of the air surrounding the tank.
User entry
0.0 to 500.0 kg/m³
Factory setting
1.2 kg/m³
Additional information
Read access
Operator
Write access
Maintenance
Measured upper density
Navigation
Operation → Density → Meas upper dens.
Description
Shows the density of the upper phase.
Additional information
Read access
Operator
Write access
-
Measured middle density
Navigation
Operation → Density → Meas middle dens
Description
Density of the middle phase.
Additional information
Read access
Operator
Write access
-
Measured lower density
Navigation
Operation → Density → Meas lower dens.
Description
Density of the lower phase.