Master Meter System
Diagnostics and troubleshooting
Hauser
51
Diagnosis
no.
Short text
Diagnosti
c
behavior
Possible cause
Remedy
255
Flow Computer 2 DUT
Temperature Low
Lower DUT temperature limit has been
exceeded.
Check process conditions.
Check settings for alarm limits.
256
Flow Computer 2 MM
Pressure High
Upper MM pressure limit has been exceeded.
Check process conditions.
Check settings for alarm limits.
257
Flow Computer 2 MM
Pressure Low
Lower MM pressure limit has been exceeded.
Check process conditions.
Check settings for alarm limits.
258
Flow Computer 2 MM
Temperature High
Upper MM temperature limit has been
exceeded.
Check process conditions.
Check settings for alarm limits.
259
Flow Computer 2 MM
Temperature Low
Lower MM temperature limit has been
exceeded.
Check process conditions.
Check settings for alarm limits.
260
Flow Computer 2 prove
aborted
An ongoing prove was aborted (either manually
or automatically).
Check detailed messages for prove. A message
will appear as soon as a new prove is started.
261
Flow Computer 2
Printer 1 Error
An error has occurred in a connected printer.
Check printer connection and settings.
262
Flow Computer 2
Printer 2 Error
An error has occurred in a connected printer.
Check printer connection and settings.
263
Flow Computer 2
Printer 3 Error
An error has occurred in a connected printer.
Check printer connection and settings.
264
Flow Computer 2
Printer Spool Full
Flow computer printer spool is full.
Check connected printers.
265
Flow Computer 2 FTP
Error
The flow computer could not save the data to
the HMI computer.
Check FTP settings on flow computer and HMI.
Contact Hauser.
266
Flow Computer 2 Duty
Meter
Communications Fail
Modbus RTU communication between the flow
computer and Promass Duty Meter has failed.
Check wiring of Modbus RTU (RS-485) cables.
Ensure that Promass is switched on. Check
Modbus RTU communication settings. Ensure
that a compatible Duty Meter is being used.
9.2
Troubleshooting
This section explains the actions the user should take to fix common computer problems
caused by hardware or software errors.
Should a problem occur, the initial steps listed below should be followed before taking
further action:
• On the tablet PC, try to identify and isolate the component that is causing the problem.
• Ensure that all peripheral devices are switched on before switching on the tablet PC.
• In the event of problems with an external device, ensure that the cable connections are
correct and secure.
• Ensure that the correct configuration information is configured in the BIOS setup
program.
• Ensure that all device drivers are correctly installed.
• Take note of the user's observations. Are there any message on the display? Are
indicator lamps lit? Are there any beeping sounds? If the user wants to seek assistance,
detailed descriptions are helpful for service staff.
If the problem persists after the user has followed the instructions in this section, contact
your local Hauser Sales Center.