Master Meter System
Diagnostics and troubleshooting
Hauser
49
Diagnosis
no.
Short text
Diagnosti
c
behavior
Possible cause
Remedy
112
Flow Computer 1 RAM
Fail
An internal flow computer error has occurred.
Restart flow computer.
If error persists, contact Hauser.
113
Flow Computer 1 RTC
Error
An internal flow computer error has occurred.
Restart flow computer.
If error persists, contact Hauser.
114
Flow Computer 1 SD
Card Error
An internal flow computer error has occurred.
Restart flow computer.
If error persists, contact Hauser.
115
Flow Computer 1 Task
Fail
An internal flow computer error has occurred.
Restart flow computer.
If error persists, contact Hauser.
116
Flow Computer 1 Units
Mismatch Error
Settings selected for units are not plausible.
Check settings for units.
117
Flow Computer 1 Flow
Rate/Total Discrepancy
Error
The pulse input signal of the MM does not
match the Modbus mass rate signal.
Check pulse input of MM and Modbus RTU
communication.
Check settings for flow discrepancy.
150
Flow Computer 1 DUT
Density High
Upper DUT density limit has been exceeded.
Check process conditions.
Check settings for alarm limits.
151
Flow Computer 1 DUT
Density Low
Lower DUT density limit has been exceeded.
Check process conditions.
Check settings for alarm limits.
152
Flow Computer 1 DUT
Pressure High
Upper DUT pressure limit has been exceeded.
Check process conditions.
Check settings for alarm limits.
153
Flow Computer 1 DUT
Pressure Low
Lower DUT pressure limit has been exceeded.
Check process conditions.
Check settings for alarm limits.
154
Flow Computer 1 DUT
Temperature High
Upper DUT temperature limit has been
exceeded.
Check process conditions.
Check settings for alarm limits.
155
Flow Computer 1 DUT
Temperature Low
Lower DUT temperature limit has been
exceeded.
Check process conditions.
Check settings for alarm limits.
156
Flow Computer 1 MM
Pressure High
Upper MM pressure limit has been exceeded.
Check process conditions.
Check settings for alarm limits.
157
Flow Computer 1 MM
Pressure Low
Lower MM pressure limit has been exceeded.
Check process conditions.
Check settings for alarm limits.
158
Flow Computer 1 MM
Temperature High
Upper MM temperature limit has been
exceeded.
Check process conditions.
Check settings for alarm limits.
159
Flow Computer 1 MM
Temperature Low
Lower MM temperature limit has been
exceeded.
Check process conditions.
Check settings for alarm limits.
160
Flow Computer 1 prove
aborted
An ongoing prove was aborted (either manually
or automatically).
Check detailed messages for prove. A message
will appear as soon as a new prove is started.
161
Flow Computer 1
Printer 1 Error
An error has occurred in a connected printer.
Check printer connection and settings.
162
Flow Computer 1
Printer 2 Error
An error has occurred in a connected printer.
Check printer connection and settings.
163
Flow Computer 1
Printer 3 Error
An error has occurred in a connected printer.
Check printer connection and settings.
164
Flow Computer 1
Printer Spool Full
Flow computer printer spool is full.
Check connected printers.
165
Flow Computer 1 FTP
Error
The flow computer could not save the data to
the HMI computer.
Check FTP settings on flow computer and HMI.
Contact Hauser.
166
Flow Computer 1 Duty
Meter
Communications Fail
Modbus RTU communication between the flow
computer and Promass Duty Meter has failed.
Check wiring of Modbus RTU (RS-485) cables.
Ensure that Promass is switched on. Check
Modbus RTU communication settings. Ensure
that a compatible Duty Meter is being used.
200
Flow Computer 2
Calculation Error
A calculation error has occurred in the flow
computer.
Restart flow computer.
If error persists, contact Hauser.