Diagnostics and troubleshooting
Liquitrend QMW43
22
Hauser
11.4.2 Overview of diagnostic events
A: Status signal/diagnostic event
B: Diagnostic behavior
C: IO-Link/EventQualifier
D: EventCode
E: Event text
A
B
C
D
E
C485
Warning
IO-Link
Warning
0x8C01
Simulation active
S825
Warning
IO-Link
Warning
0x1812
Ambient temperature outside of
specification
S971
Warning
IO-Link
Warning
0x1811
Measured value is outside sensor
range
F270
Fault
IO-Link
Error
0x5000
Defect in electronics/sensor
S803
Fault
IO-Link
Error
0x1804
Current loop
S804
Fault
IO-Link
Error
0x1801
Overload at output 1/2
C103
Message
IO-Link
Message
0x1813
Sensor check failed
C182
Message
IO-Link
Message
0x1807
Invalid calibration
-
Message
IO-Link
Message
0x1814
Sensor check passed
1)
EventCode as per IO-Link standard 1.1
Causes and remedial action
Warnings
C485
When the simulation of conductivity or buildup is active, the device displays a warning.
‣
Deactivate simulation.
S825
Ambient temperature outside of specification.
‣
Operate the device in the specified temperature range.
S971
The measured value is outside the set sensor range.
‣
Operate the device in the configured measuring range or set the measuring range.
Faults
F270
Electronics/sensor defective.
‣
Replace device.
S803
Impedance of load resistance at analog output too high.
‣
Check the cable and load at the current output.
‣
If a current output is not required, then switch off the current output via the
configuration.
‣
Connect current output to load.