Commissioning
Liquitrend QMW43
16
Hauser
Identifier
ISDU (dec) ISDU (hex)
Size [byte] Data type Access Default value
Data storage
VendorText
17
0x0011
max. 64
String
r/-
People for Process Automation
Device ID
9 to 11
0x0009 to 0x000B
r/-
0x000600
Hardware Revision
22
0x0016
max. 64
String
r/-
Application Specific Tag 24
0x0018
32
String
r/w
Actual Diagnostics (STA) 260
0x0104
4
String
r/-
No
Last Diagnostic (LST)
261
0x0105
4
String
r/-
No
8.2.3
System commands
Identifier
ISDU (dec)
ISDU (hex)
Value range
Access
Reset to factory settings (RES)
130
0x0082
-/w
Device Access Locks.Data Storage Lock
12
0x000C
0 ~ False
2 ~ True
9
Commissioning
9.1
Function check
Prior to commissioning, make sure that the post-installation and post-connection checks
have been performed.
Checklists in sections
• Post-installation check
• Post-connection check
9.2
Light signals (LEDs)
1
2
2
1
3
4
A0041157
9
Position of LEDs in housing cover
1
green (GN), status, communication
2
red (RD), warning or fault
Description of the function of the LEDs
Position 1: green (GN) status, communication
• Lit: no communication
• Flashing: active communication, flash frequency
• Flashing with increased luminosity: device search (device identification), flash frequency