System integration
Liquitrend QMW43
14
Hauser
7.3
Structure of the operating menu
"Overview of the operating menu" section → 28
8
System integration
8.1
Process data
Bit
0 (LSB)
1
...
22
23 (MSB)
24
...
31
Device
Conductivity: 0 to 110
000 µS/cm, resolution 0.1 µS/cm
Buildup: 0 to 10 mm, resolution 0.1 mm
UInt24: offset = 0, gradient = 0.1
UInt8: offset = 0, gradient = 0.1
In addition, the value for buildup and conductivity can be read via ISDU (hex) 0x0028
– acyclic service.
8.2
Reading out and writing device data (ISDU – Indexed
Service Data Unit)
Device data are always exchanged acyclically and at the request of the IO-Link master.
Using the device data, the following parameter values or device statuses can be read out:
8.2.1
Hauser-specific device data
Identifier
ISDU
(dec)
ISDU
(hex)
Size
[byte]
Data type
Access Default value
Value range
Offset/
gradient
Data
storage
Range limits
Extended order
code
259
0x0103
60
String
r/-
ENP_VERSION
257
0x0101
16
String
r/-
Device Type
256
0x0100
2
Uinteger16 r/-
0x91FB
Simulation
buildup
66
0x0042
1
UInt8
r/w
0 ~ OFF
0 ~ OFF
1 ~ ON
No
0 to 1
Simulated
buildup
85
0x0055
4
Int16
r/w
10
0 to 10.0
Yes
0 to 10.0
Simulation
conductivity
90
0x005A 1
UInt8
r/w
0 ~ OFF
0 ~ OFF
1 ~ ON
No
0 to 1
Simulated
conductivity
86
0x0056
4
UInt32
r/w
100
000.0
0 to 110
000.0
Yes
0 to 110
000.0
Device search
69
0x0045
1
UInt8
r/w
0 ~ OFF
0 ~ OFF
1 ~ ON
0/1
No
0 to 1
Sensor check
70
0x0046
1
UInt8
-/w
0 ~ OFF
0 ~ OFF
1 ~ ON
0/1
No
0 to 1
Operating Mode
1 (OU1)
108
0x006C
1
UInt8
r/w
OFF
0 ~ OFF
3 ~ Frequency
Yes
0 to 4
Operating Mode
2 (OU2)
97
0x0061
1
UInt8
r/w
3~ Frequency
(Conductivity)
2 ~ 4-20 mA
(Buildup)
3 ~ Frequency
4 ~ 4-20 mA
(Conductivity)
Yes
0 to 4