Description of Device Parameters
Liquitrend QMW43
34
Hauser
Note
The configured measuring range for buildup is always = 0 to 10. The parameter is
therefore not shown/displayed in this case. The configured measuring range for
conductivity can be changed.
The span can be ordered as a preset value in the product structure via order code 570
"Service", option HT "Configuration output 1 + output 2".
Description
For specifying the measuring range for the frequency output.
Input range
Minimum span
0 to 110
000.0
1000,0
Switch on value
Last configured value.
16.3.2 System
Operating hours
Navigation
Parameter → System → Operating hours
Description
This parameter counts the operating hours during the period in which operating voltage is
present. The value is output in industrial hours.
µC-temperature
Navigation
Parameter → System → µC-temperature
Description
This parameter displays the current µC-temperature on the electronics.
Unit changeover (UNI) - µC-Temperature
Navigation
Parameter → System → Unit changeover (UNI) - µC-Temperature
Description
This parameter is used to select the electronics temperature unit. Once a new electronics
temperature unit has been selected, the value is converted to the new unit and displayed.
Switch on value
Last unit selected prior to switching off.
Options
°C
°F
K
Factory setting
°C
Minimum µC-Temperature