Commissioning
Proline Promass 83
70
Hauser
6.4.2
Current input: active/passive
The current outputs are configured as "active" or "passive" by means of various jumpers on the
current input submodule.
#
Warning!
Risk of electric shock.
Exposed components carry dangerous voltages. Make sure that the power supply is switched
off before you remove the cover of the electronics compartment.
1.
Switch off power supply
2.
3.
Set the jumpers
"
Caution!
– Risk of destroying the measuring device. Set the jumpers exactly as shown in the
diagram. Incorrectly set jumpers can cause overcurrents that would destroy either the
measuring device or external devices connected to it.
– Note that the position of the current submodule on the I/O board can vary, depending
on the version ordered, and that the terminal assignment in the connection
compartment of the transmitter varies accordingly
28.
4.
Installation of the I/O board is the reverse of the removal procedure.
a0005124
Fig. 41:
Configuring current inputs with the aid of jumpers (I/O board)
Current input 1 (optional, plug-in module)
1
Active current input (default)
2
Passive current input
INPUT/OUTPUT
4
1
2
+
+
INPUT/OUTPUT
3
INPUT/OUTPUT
2