Proline Promass 83
Technical data
Hauser
107
Measuring ranges for gases, general, (except Promass H (Zr))
The full scale values depend on the density of the gas. Use the formula below to calculate the
full scale values:
g
max(G)
=
g
max(F)
× r
(G)
: x [kg/m
3
(lb/ft
3
)]
g
max(G)
= Max. full scale value for gas [kg/h (lb/min)]
g
max(F)
= Max. full scale value for liquid [kg/h (lb/min)]
r
(G)
= Gas density in [kg/m
3
(lb/ft
3
)] for process conditions
Here,
g
max(G)
can never be greater than
g
max(F)
Measuring ranges for gases (Promass F, O):
Measuring ranges for gases (Promass E)
Measuring ranges for gases (Promass P, S, H (Ta))
DN
x
[mm]
[inch]
8
3/8
60
15
½
80
25
1
90
40
1½
90
50
2
90
80
3
110
100
4
130
150
6
200
250
10
200
DN
x
[mm]
[inch]
8
3/8
85
15
½
110
25
1
125
40
1½
125
50
2
125
80
3
155
DN
x
[mm]
[inch]
8
3/8
60
15
½
80
25
1
90
40
1)
1½
1)
90
50
1)
2
1)
90
1)
only Promass P, S