Technical data
Proline Promass 83
106
Hauser
10
Technical data
10.1
Applications
10.2
Function and system design
Measuring principle
Mass flow measurement by the Coriolis principle
Measuring system
10.3
Input
Measured variable
• Mass flow (proportional to the phase difference between two sensors mounted on the
measuring tube to register a phase shift in the oscillation)
• Fluid density (proportional to resonance frequency of the measuring tube)
• Fluid temperature (measured with temperature sensors)
Measuring range
Measuring ranges for liquids
DN
Range for full scale values (liquids)
g
min(F)
to
g
max(F)
[mm]
[inch]
[kg/h]
[lb/min]
1
1/24
0 to 20
0 to 0.735
2
1/12
0 to 100
0 to 3.675
4
1/8
0 to 450
0 to 16.54
8
3/8
0 to 2000
0 to 73.50
15
½
0 to 6500
0 to 238.9
15 FB
½ FB
0 to 18000
0 to 661.5
25
1
0 to 18000
0 to 661.5
25 FB
1 FB
0 to 45000
0 to 1654
40
1 ½
0 to 45000
0 to 1654
40 FB
1 ½ FB
0 to 70000
0 to 2573
50
2
0 to 70000
0 to 2573
50 FB
2 FB
0 to 180000
0 to 6615
80
3
0 to 180000
0 to 6615
100
4
0 to 350000
0 to 12860
150
6
0 to 800000
0 to 29400
250
10
0 to 2200000
0 to 80850
350
14
0 to 4100000
0 to 150675
FB = Full bore versions of Promass I