CleanFit P CPA 472
3 Installation
16
Hauser
3.3.5 Installing the assembly
Set the assembly in maintenance position (see Fig. 3.7) and attach it to the selected
connection on the tank or pipe.
Process connections are available for the CPA 472 assembly:
G 1¼ external thread, Tri-Clamp 2", flange DN 50, flange ANSI 2", NPT 1" external
thread, grooved nut Rd 78
Abb. 3.15: Existing process connections for the CleanFit P CPA 472 assembly
The immersion depth is specified in each case (long assembly version in brackets)
!
Hinweis!
Note the following depending on the process connection fitted:
All connections:
•
Check the correct seating of the O-rings on the assembly.
Abb. 3.16: O-ring seals of the CPA 472 (in contact with medium)
Flange connection:
Check the seat of the flange seal between the flanges.