CleanFit P CPA 472
7 Technical Data
Hauser
25
7
Technical Data
General specifications
Manufacturer
Hauser
Product name
CleanFit P CPA 472
Ambient conditions
Ambient temperature (nominal operating conditions)
> 0°C!
Process conditions
Process temperature range
0 ... 90°C (dependent on selected material and pro-
cess pressure)
Process pressure range
0 ... 4 bar in manual mode
0 ... 6 bar in pneumatic mode
Physical data
Electrode* immersion depth
up to 110 mm (short version)
up to 215 mm (long version)
Required freedom of installation
approx. 0.5 m
Electrode lengths
120 mm, 225 mm
Weight of assembly
approx. 1.5 kg (short version)
approx. 3 kg (long version)
Materials
Materials in contact with medium
Electrode holder PP, PEEK, PVDF
Seals: VITON
®
, KALREZ
®
Materials in contact with medium
Housings PA
Process connections
G 1¼ external thread, flange DN 50, flange ANSI 2", NPT 1" external thread, grooved nut Rd 78
Network connections
Rinse connections
2 x G ¼ (ID)
2 x NPT ¼" (ID)
Rinse water pressure 2 ... 6 bar
Compressed air connections
Pressure 4 ... 8 bar
air filtered (40 µm), water- and oil-free
air tubes with minimum DN 4 mm
Limit switches
pneumatic 3/2-way valve
* At an immersion depth of 215 mm no CeraLiquid P electrodes (CPS 41/42) can be installed.